MAINTENANCE MENU: ERROR CODES

Scale Error Codes
16 = Angle sensor defective .
17 = Scale chip failure . The scale chip on the scale board is faulty .
For errors 16 and 17, unplug the bed and wait 5 seconds . Then plug the bed back in . If the error persists, replace the
scale board and calibrate the bed scale .
18 = Value of the foot left load cell is out of range: load cell may be disconnected, faulty or scale board is faulty .
19 = Value of the head right load cell is out of range: load cell may be disconnected, faulty or scale board is faulty .
20 = Value of the foot right load cell is out of range: load cell may be disconnected, faulty or scale board is faulty .
21 = Value of the head left load cell is out of range: load cell may be disconnected, faulty or scale board is faulty .
For errors 18 to 21, first verify load cell cables are plugged in and inspect for broken wires . If no visible problem is found,
swap two load cells on the scale connectors . This will help determine whether the load cell or the scale board is faulty .
If the error code changes with the load cell, the problem is with the load cell . Replace the load cell . However, if the error
code remains the same, then the problem is with the scale board . Replace scale board . In both cases, the bed has to
be calibrated .