Intended Use Of Product; Scale System - Stryker GoBed II Lit MedSurg FL28EX Guide D'utilisation

Lit d'hôpital ajustable
Masquer les pouces Voir aussi pour GoBed II Lit MedSurg FL28EX:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

This manual is designed to assist you with the operation of the GoBed
this manual thoroughly before using the equipment or beginning maintenance on it . To ensure safe operation of this
equipment, it is recommended that methods and procedures be established for educating and training staff on the safe
operation of this bed .

INTENDED USE OF PRODUCT

®
The GoBed
II MedSurg Bed, Model FL28EX is an AC-powered adjustable hospital bed designed to position patients for
procedures, therapy, and recovery in healthcare environment, transport patients between bays and procedural rooms,
provide patient security, measure and display patient weight, allow patients to alert caregiver when the patient requires
emergency assistance or any assistance to improve comfort levels, and monitor patient position to alert caregiver of
potential fall .
Intended operators are health care professionals (nurses, nurse aids, doctors) which can use all bed operations (e .g .,
bed motion functions, nurse call, siderail communications, bed exit), patient and bystander, which can use bed motion
functions, nurse call & siderail communications, and trained professionals for installation, service, and calibration .
The product is intended to be used in a healthcare environment, including hospitals, surgery centers, long term acute
care centers, and rehabilitation centers . The intended environmental conditions are 10 to 40 degrees Celsius and 30
to 75% RH . The product is compatible with 35" x 84" surfaces, the facility nurse call system, standard Med-Surg
equipment, and the facility infrastructure . Its expected life is 10 years .
Intended patient population are general hospital and acute care patients . The bed accommodates a safe working load of
500 pounds . Patients must weigh more than 50 pounds to use the bed exit system . It is not intended for pediatric use
and/or patients measuring 35" or less .
SPECIFICATIONS
Safe Working Load
Note: Safe Working Load indicates the sum of
the patient, mattress, and accessory weight .
Scale System (Optional)
- Accuracy of scale display*
- Operating Angular Range
* To meet this accuracy claim, the patient surface must be in the flat position (fowler and gatch down) and the
stretcher cannot exceed 5 degrees of Trendelenburg/reverse Trendelenburg .
Overall Length/Width
- Siderails Up
- Siderails Down
Weight with Headboard and Footboard
Patient Sleep Surface
Recommended Mattress Size
Mattress Maximum Thickness
Minimum/Maximum Bed Height
Fowler Angle
www.stryker.com
Introduction
®
II MedSurg Bed, Model FL28EX . Carefully read
500 lb
± 2% of the total patient weight for patient weighing
greater than 100 pounds (45,3 kg)
± 2lbs below 100 lbs (45,3 kg)
-12° to +12°
94 5/8" x 40" (240,34 cm x 101,6 cm)
94 5/8" x 39" (240,34 cm x 99,06 cm)
472 lb (214,1 kg)
35 x 84" (89 x 213 cm)
35 x 84" (89 x 213 cm)
6" (15,24 cm)
14 1/2" to 29" (36,8 to 73,7 cm)
0° to 59°
28EX-009-001 REV B
227 kg
Return To Table of Contents
1-5

Publicité

Chapitres

Table des Matières
loading

Table des Matières