GATCH MOTOR REPLACEMENT (CONTINUED)
17 . To adjust the limits of the replacement thigh motor, proceed as follows:
A . Ensure the motor cable is connected to the control board and plug the power cord into the wall receptacle .
B . Position the new actuator tube holes horizontally . Grasp the tube firmly to prevent it from rotating, and press
the Gatch up control for a few seconds . Then press the down control until the motor stops . This will be the
lower limit of the motor .
C . Gently turn the tube in either direction to align the tube holes with those of the thigh section lever arms . Then
raise the Gatch a few inches while firmly holding the tube and lower it completely .
D . Check the alignment of the holes . If the tube holes are not aligned with those of the lever arms, repeat steps
B and C until they are . Once the holes are aligned, install the washers and the clevis pin . Slightly raise the
thigh section manually to ease the insertion of the clevis pin .
E . Before installing the rue ring cotter, raise and lower the Gatch completely to ensure the motor stops when
the thigh section reaches the frame .
18 . Test the Gatch section for proper operation before returning the bed to service .