WARNING
Always unplug the bed power cord from the wall outlet when servicing or cleaning the bed .
Hand wash all surfaces of the
thoroughly . Do not steam clean or hose off the bed . Using these methods of cleaning is not recommended and may
void this product's warranty . Do not immerse any part of the bed in any kind of liquid . Some of the internal parts of the
bed are electric and may be damaged by exposure to water .
Suggested cleaners for bed surfaces:
• Quaternary Cleaners (active ingredient - ammonium chloride) .
• Phenolic Cleaners (active ingredient - o-phenylphenol) .
• Chlorinated Bleach Solution (5 .25% - less than 1 part bleach to 100 parts water) .
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer's guidelines for
proper disinfecting .
CAUTION
Do not steam clean, hose off or ultrasonically clean the bed . Do not immerse any part of the bed in any kind of liquid .
The internal electrical parts may be damaged by exposure to water . Hand wash regularly all surfaces of the bed with
warm water and a mild detergent . Wipe cleaned surfaces dry to avoid build up of cleaning substance . Inspect the
mattress after each use . Discontinue use if any cracks or rips, which may allow fluid to enter the mattress, are found in
the mattress cover . Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of exposure
to pathogenic substances and may bring about diseases to the patient and user .
CAUTION
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED
IMPROPERLY . If the products suggested above are used to clean Stryker patient care equipment, measures must be
taken to insure the bed is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning . Failure
to properly rinse and dry the bed will leave a corrosive residue on the surface of the bed, possibly causing premature
corrosion of critical components . Failure to follow the above directions when using these types of cleaners may void this
product's warranty .
For mattress cleaning instructions, please see the tag on the mattress, or contact the mattress manufacturer .
Clean Velcro
®
after each use . Saturate Velcro
disinfectant for nylon Velcro
®
Return To Table of Contents
1-14
Cleaning
®
GoBed
II MedSurg Bed, Model FL28C
®
with disinfectant and allow disinfectant to evaporate . (Appropriate
should be determined by the hospital) .
280C-009-002 REV B
with warm water and mild detergent . Dry
www.stryker.com