II MEDSURG BED, MODEL FL28C (CONTINUED)
The GoBed
II MedSurg Bed, Model FL28C is suited for use in the electromagnetic environment specified below .
®
The customer or the user of the GoBed
environment .
Immunity Test
IEC 60601 Test Level
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
80 MHz to 2 .5 GHz
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies .
Note 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy .
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered . If the measured field strength in the location in which the GoBed
exceeds the applicable RF compliance level above, the GoBed
to verify normal operation . If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the GoBed
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m .
Return To Table of Contents
1-164
EMC Information
®
II MedSurg Bed, Model FL28C should assure that it is used in such an
Compliance Level
3 Vrms
3 Vrms
3 V/m
3 V/m
®
II MedSurg Bed, Model FL28C .
280C-009-002 REV B
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the GoBed
®
II MedSurg Bed, Model FL28C,
including cables, than the recommended
separation distance calculated from the
equation appropriate for the frequency of the
transmitter .
Recommended Separation Distance
d=1 .2
d=1 .2
d=2 .3
800 MHz to 2 .5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m) .
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range .
Interference may occur in the vicinity of
equipment marked with the following symbol:
®
II MedSurg Bed, Model FL28C is used
®
II MedSurg Bed, Model FL28C should be observed
a
b
www.stryker.com