HEADWALL CONTROL BOARD REPLACEMENT (OPTIONAL)

Tools Required:
• #2 Phillips Screwdriver
C
D
B
Procedure:
1 .
Raise the bed to the highest position and apply the
brakes .
2 .
Unplug the bed power cord from the wall outlet .
3 .
Remove the power cord from the power connector and
disengage it from the wire clip .
4 .
Properly ground yourself (see "Grounding Diagram"
page
1-13) .
5 .
Using a #2 Phillips screwdriver, remove the six screws
(A) holding the cover to the litter . Keep the cable clips .
6 .
Once the screws removed, pivot the cover and lay it flat
on the bottom of the litter (see opposite illustration) .
7 .
Remove all cables connected to the control board . Note
their location so they will be connected properly to the new board . Refer to drawing OL280120,
OL280121,
page
1-94, for the connecting position of the cables on the power connector .
8 .
Turn the cover to a vertical position and . Using a #2 Phillips screwdriver, remove the four screws (B) holding the
control board (C) to the head casing cover and remove the board .
9 .
Reverse the above steps to install the new control board . Before installing the new board, ensure the dip switch
settings on the new board match the ones on the old board .
10 . Test the nurse call for proper operation before returning the bed to service .