FOWLER REPLACEMENT (CONTINUED)
11 . Using two 1/2" combination wrenches, remove the locknut and bolt (M, Figure 17,
cylinder end fitting to the head section lever arms . Lay the CPR mechanism down on the frame .
12 . Grasp the head section firmly . Remove the strap and lower the head section .
13 . Using a 1/2" combination wrench, remove the two locknuts, two shoulder spacers and two bolts (F) holding the head
section to the seat section .
14 . Remove the micro switch from the defective head section by pressing its two side clips .
15 . Reverse the above steps to install the new head section .
16 . Test the head section and CPR mechanism for proper operation before returning the bed to service .
17 . Using two 1/2" combination wrenches, remove the locknut and bolt (A, Figure 7) holding the lever to the head
section . Lay the lever down on the frame . Keep the spacer (B, Figure 7) .
18 . Grasp the head section firmly . Remove the strap and lower the head section .
19 . Using a 1/2" combination wrench, remove the two locknuts, two shoulder spacers and two bolts (F) holding the head
section to the seat section .
20 . Reverse the above steps to install the new head section .
21 . Test the head section for proper operation before returning the bed to service .
www.stryker.com
Service Information
Figure 7
280C-009-002 REV B
page
1-59) holding the pneumatic
A
B
Return To Table of Contents
1-39