Stryker GoBed II MedSurg Bed Guide D'entretien page 165

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

®
GOBED
II MEDSURG BED, MODEL FL28C (CONTINUED)
Recommended separation distances between portable and mobile RF
communications equipment and the GoBed
®
The GoBed
II MedSurg Bed, Model FL28C is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled . The customer or the user of the GoBed
FL28C can help prevent electromagnetic interferences by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the GoBed
recommended below, according to the maximum output power of the communications equipment .
Rated maximum output
power of transmitter
W
0 .01
0 .1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer .
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
Note 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
www.stryker.com
EMC Information
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d=1 .2
1 .12
0 .38
1 .2
3 .8
12
280C-009-002 REV B
®
II MedSurg Bed, Model FL28C.
®
II MedSurg Bed, Model
®
II MedSurg Bed, Model FL28C as
m
800 MHz to 2 .5 GHz
d=1 .2
0 .12
0 .38
1 .2
3 .8
12
d=2 .3
0 .23
0 .73
2 .3
7 .3
23
Return To Table of Contents
1-163

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Gobed ii lit medsurgFl28c

Table des Matières