LOAD CELL REPLACEMENT (OPTIONAL) (CONTINUED)
Support
Point
6 .
Using diagonal pliers, cut the two cable ties holding the defective cell cable to the frame immediately after the cell .
Repeat for the adjacent cell .
7 .
Lower the bed until the two Allen screws holding the defective load cells to the frame are easily accessible .
8 .
Unplug the power cord from the wall receptacle .
9 .
Using a 1/4" Allen wrench and a 1/2" combination wrench, remove the two nuts and two Allen screws (B) holding
the defective load cell to the frame .
10 . Cut the cable ties holding the defective load cell cable the frame . Unplug the defective load cell cable from the
scale/bed exit board . Remove the defective load cell .
11 . Inspect the elastomer sleeves (C) to ensure that the inner circular hole is not worn . Replace if necessary . Note the
orientation of the elastomer sleeves when replacing (see detail at right) .
12 . Reverse the above steps to install the new load cell .
13 . Calibrate the scale before returning the bed to service . Refer to the "Scale
Calibration" procedure,
page
Return To Table of Contents
1-54
Service Information
1-55 .
280C-009-002 REV B
Support
Point
Proper elastomer
sleeve orientation.
www.stryker.com