HEAD END LIFT MOTOR REPLACEMENT (CONTINUED)
12 . To adjust the limits of the new motor, proceed as follows:
A . Ensure the motor cable is connected to the control board and plug the power cord into the wall receptacle .
B . Press the bed down control until the motor stops . This will be the lower limit of the motor .
C . Attach the molded nut support on the motor to the lift lever arms .
D . With the alignment fixtures still in place, raise and lower the bed completely to ensure the lower limit is
properly adjusted .
13 . Test the Bed Up and Bed Down functions for proper operation before returning the bed to service .

FOOT END LIFT MOTOR REPLACEMENT

Note
In order to maintain the adjustment of the low position when replacing a lift motor, a special tool kit must be used . The kit
includes alignment fixtures . To obtain this kit, contact our Customer Service department and order part number KR0154 .
Tools Required:
• Tool Kit KR0154
• Cutting Pliers
• #2 Phillips Screwdriver
• Small Regular Screwdriver
• 1/2" Combination Wrench
• Bungee Cord (or equivalent)
Note
When raising or lowering any of the litter sections, ensure that all siderails are in the highest position . Unless otherwise
indicated, refer to Figure 10,
page
1 .
Install the alignment fixtures (KR0154) on both sides of the brake mechanism on the base frame, under the lower
fixed frame .
2 .
Lower the bed, until the lower fixed frame is completely supported by the U-shaped pieces of the fixture (see Figure
9,
page
1-44) .
3 .
Remove the mattress and the plastic litter cover (if equipped) from the foot section metal frame .
4 .
Raise the fowler section to about 45°, and raise the thigh section to the highest position .
5 .
Fold the foot section back toward the head end of the bed . Secure it using a bungee cord .
6 .
Unplug the power cord from the wall receptacle .
7 .
Using a #2 Phillips screwdriver, remove the four screws holding the frame cover plate .
8 .
Using cutting pliers, cut the cable ties holding the defective lift motor power cord to the bed frame . Unplug the motor
connector .
9 .
Using a 1/2¨ combination wrench, remove the two bolts (D), two washers (C) and two spacers (B) holding the
molded nut support to the elevation lever arm .
10 . Using a #2 Phillips screwdriver, remove the two screws holding the motor support cover to the lift lever arm and
remove it .
11 . Remove the two pivot pins (H) holding the motor (A) to the support . To ease the removal of the pins, insert a small
regular screwdriver into the opening at the end of the motor and push out the pins .
12 . Remove the defective motor by passing it under the bed frame .
13 . Using a #2 Phillips screwdriver, remove the two screws (F) holding the plastic screw cover (J) to the motor . Remove
the plastic screw cover and install it on the new actuator .
www.stryker.com
Service Information
1-44, for reference for this procedure .
280C-009-002 REV B
Return To Table of Contents