WARNING (CONTINUED)
• To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the steer mode engaged .
The steer wheel cannot swivel .
• The CPR emergency release is for emergency use only . When activating the CPR release handle, all people and
equipment must be removed from the area below and around the head, thigh and foot sections of the bed or serious
personal injury and/or equipment damage could occur .
• Possible fire hazard exists when this unit is used with oxygen administering equipment other than nasal, mask type
or half bed-length tent type . Unplug the unit power cord from the wall when oxygen-administering equipment is used .
When using a half bed-length tent type, ensure that the siderails are outside the oxygen tent and that the tent does
not extend below the mattress support level .
• The weight system (optional) is intended to assist in the monitoring of the patient's weight variation . Under no
circumstances should its readings be used as sole reference for medical treatment .
• The Bed Exit system (optional) is intended only to aid in the detection of a patient exiting the bed . It is not intended to
replace patient monitoring protocol . The Bed Exit system signals when a patient is about to exit the bed . The addition
or removal of equipment with a Bed Exit system armed must be done using the "Adding or Removing Equipment
with the System Armed" procedure, otherwise the sensitivity of the system may be affected and the readings of the
patient's movements in the bed will be erroneous .
• The surface added by the bed extender is not included in the area monitored by the Bed Exit system .
• The Bed Exit system (optional) is not designed to be used with patients weighing less than 50 lb (23 kg) .
• Failure to set the zone properly could result in improper monitoring and patient injury . Verify that the proper zone is
selected before leaving the patient .
• The scale and bed exit systems must be adjusted if an accessory is added when either system is in operation . Refer
to the "Add / Remove Equipment" procedure .
• The bed extender must be removed before using the emergency crank (optional) .
• When large fluid spills occur in the area of the circuit boards, cables and motors, immediately unplug the bed power
cord from the wall outlet, remove the patient from the bed and clean up the fluid . Have maintenance completely
check the bed . Fluids can have an adverse effect on operational capabilities of any electrical product . Do not put
the bed back into service until it is completely dried and has been thoroughly tested for safe operation . Ensure,
among other things, that the plastic components being used as covers for the siderail mechanism arms and the foot
end casing are removed and that the parts they cover are thoroughly dried .
• Preventative maintenance should be performed at least once a year to ensure all bed features are functioning
properly . Ensure that any bed malfunction is promptly reported to your service personnel for immediate attention .
• Always unplug the bed power cord from the wall outlet when servicing or cleaning the bed .
• When working under the bed with the bed in the high position, always apply the brakes and place blocks under the
Hi-Low levers to prevent injury in case the bed down switch is accidentally pressed .
• Before using the optional emergency crank during a power failure, always unplug the power cord . An unexpected
return of the power could rotate the handle and cause injury to the user .
• Because individual beds may have different options, foot boards should not be moved from one bed to another .
Mixing foot boards could result in unpredictable bed operation .
• When servicing use only identical replacement parts provided by Stryker .
• Make sure that the ground chain is in place, intact and it's touching the floor (see Figure 1 .9,
• Do not put anything under the bed .
• Only service technicians from Stryker or service personnel trained by Stryker should perform the procedures
detailed in this maintenance manual . This is especially important for procedures related to the Scale (option) and
Bed Exit (option) systems . Failure to follow this restriction can result in serious damage to the bed and/or severe
injury to users or patients .
• To prevent injury when working under the bed with the bed in the high position, always place blocks under the
mattress support frame and apply the brakes .
www.stryker.com
Safety Tips and Guidelines
280C-009-002 REV B
page
1-8) .