SCALE SYSTEM CONTROL PANEL GUIDE (OPTIONAL)
WARNING
The weight system (optional) is intended to assist in the monitoring of the patient's weight variation . Under no circumstances
should its readings be used as sole reference for medical treatment .
Note
When the bed is powered on, the scale display will show a welcome message and the version (X .X) of the control
software:
I:
/ Enter Control
• Press
once to activate the display . The Scale mode will be displayed .
• The display may be shut off by pressing the
• Press ENTER when instructed to do so .
J:
Zero Control
• To zero the scale
H:
Menu Up/Down Controls
• Press either control repeatedly until desired mode is reached .
G:
Scale Display
Scale Menu Modes
Order of appearance of the modes when the MENU Down control is pressed repeatedly:
Mode
Scale
Gain or Loss
Gain or Loss Reference Weight
Change Equipment
Change Patient Weight
Select Weight Unit (lb/kg)
Note
• The scale readings must be within the range specified in the Specifications section of this manual . Otherwise, the
scale must be calibrated . Refer to the scale calibration procedure of the Maintenance Manual .
• The scale does not operate when the bed is inclined more than 12° in the Trendelenburg or reverse Trendelenburg
position . The LCD will display a warning message and the current bed angle, but not the patient's weight .
• Should the message "No Scale Found" appear on the screen during the normal course of operation of the scale,
press the enter key . Data will reappear on the screen and the current operation will resume . If the message
remains, contact Technical Support .
www.stryker.com
Operation Guide
G
H
Figure 2.14B
S
T
R
Y
K
S
C
A
L
E
key .
First Line Appearing
Weight Angle
Gain Or Loss
Gain/loss Zero
Change Equipment
Change Patient Weight
Units
280C-009-001 REV B
I
J
E
R
V
X
.
X
Return To Table of Contents
1-19