IBP Alarm information
The IBP alarm information will be displayed on the alarm information area.
Technical alarm: "IBP1 Probe off" and so on.
Physiological alarm: "IBP1 Low SYS " and so on.
12.5 Changing IBP Settings
Select "Menu" → "IBP" to enter into IBP related setting.
Label: the name of blood pressure to be measured. It has the following options:
ART---artery pressure
PA---pulmonary pressure
CVP---central vain pressure
RAP---right atrium pressure
LAP---left atrium pressure
ICP---intracranial pressure
AUXP1---auxiliary pressure 1
AUXP2---auxiliary pressure 2
AUXP -- auxiliary pressure: AUXP1/2 can be chosen if the actual measuring pressure is not in the
list of ART, PA, CVP, RAP, LAP, and ICP. The operator can choose the Calculating Mode according to
the type of blood pressure. If artery pressure is monitored, please choose "dynamic", then the
reading out will include Systolic, Diastolic and MAP. If vein pressure is monitored, please choose
"static", then only the MAP will be displayed.
Calc. Mode: when the Label is AUXP1 or AUXP2, the calculate mode can be "static" or "dynamic".
Filter: there are two options for the pressure waveform filtering ---12.5Hz and 40Hz, the factory is
12.5Hz.
Avg. time: the time period for averaging to calculate Mean Pressure. Setting range from 1s to 12s,
the factory is 8s.
Unit: pressure unit. 2 options: mmHg and kPa.
Zero: perform zero calibration for pressure transducer. Press "Zero" button, then zero calibration
dialog pops up on the screen, then press "Zero" to start calibration. (Note: before performing zero
calibration, please make sure that the transducer is well connected, or the zero calibration will not
work.)
PARM: to enter into IBP alarm setting. See Section "Understanding the Alarm Setting".