Télécharger Imprimer la page

Gima K10 Manuel D'utilisation page 63

Moniteur multiparamètres

Publicité

Les langues disponibles

Les langues disponibles

CSI
6.4.2 Factory Default Alarm Limit Setting Value
Parameter
High limit
HR
Low limit
High limit
RR
Low limit
High limit
TEMP
Low limit
High limit
SYS
Low limit
High limit
DIA
Low limit
High limit
MAP
Low limit
High limit
SpO
2
Low limit
High limit
S-T segment
Low limit
High limit
EtCO
2
Low limit
High limit
FiCO
2
Low limit
High limit
PR
Low limit
TD
CSI
High limit
RAP
LAP
ICP
(Low limit+1) -300
AUXP1
(Low limit+1) -300
AUXP2
Type
Adult
180 bpm
40 bpm
30 rpm
8 rpm
39 ℃
35 ℃
180 mmHg
60 mmHg
120 mmHg
50 mmHg
160 mmHg
50 mmHg
100 %
90 %
+1.00mV
-1.00mV
50mmHg
25mmHg
4mmHg
180 bpm
40 bpm
2 ℃
/
60
(Low limit+1) -40
(Low limit+1) -40
(Low limit+1) -40
(Low limit+1)-100
Pediatric
200 bpm
130 mmHg
50 mmHg
90 mmHg
40 mmHg
110 mmHg
40 mmHg
+1.00mV
-1.00mV
50mmHg
25mmHg
4mmHg
0
200 bpm
51
-10 - (High limit-1)
-10 - (High limit-1)
-10 - (High limit-1)
-50 - (High limit-1)
-50 - (High limit-1)
0-(High limit-1)
50 bpm
30 rpm
8 rpm
39 ℃
35 ℃
100 %
85 %
0
50 bpm
2 ℃
60
Neonate
220 bpm
50 bpm
100 rpm
30 rpm
39 ℃
35 ℃
110 mmHg
50 mmHg
90 mmHg
30 mmHg
100 mmHg
30 mmHg
100 %
85 %
+1.00mV
-1.00mV
45mmHg
30mmHg
4mmHg
0
220 bpm
50 bpm
2 ℃
60

Publicité

loading

Ce manuel est également adapté pour:

K12K1535306353073530835309 ... Afficher tout