3.8.4 Printer Settings
Select "Menu"→"System Settings"→"Printer".
Speed: the printer sweep speed. Options are 25mm/s and 50mm/s. The default is 25mm/s.
⚫
Duration: The duration of printing real-time waveform. Options are: 10, 20, 30, 40, 50, 60
⚫
(seconds), and Continuous, the default is "10" seconds. If "Continuous" is selected, then the
printer continues to print until the manual stop it by pressing the print key.
Wave 1: print the first channel of waveform. Optional are varied according to the leadwire type.
⚫
The default is ECG II.
➢
3 leadwire: If "3-lead wires" is selected, then this item is in grey and nonadjustable. The lead is
same as that displayed on the screen. The first channel of waveform displaying on the current ECG
window will be the first waveform that the device will print (it can be set by Menu→ECG→Lead).
➢
5 leadwire: "ECG I", "ECG II", "ECG III", "ECG aVR", "ECG aVL", "ECG aVF" and "ECG V" (that's ECG
lead I~V).
➢
10 leadwire: "ECG I", "ECG II", "ECG III", "ECG aVR", "ECG aVL", "ECG aVF", "ECG V1", "ECG V2",
"ECG V3", "ECG V4", "ECG V5" and "ECG V6".
Wave 2: print the second channel of waveform. If "5-lead wires" is selected, then the optional will
⚫
be: OFF, "ECG I", "ECG II", "ECG III", "ECG aVR", "ECG aVL", "ECG aVF", "ECG V", "SpO
"3-lead wires" is selected, then the optional will be: "OFF", "SpO
selected, the option will be OFF, "ECG I", "ECG II", "ECG III", "ECG aVR", "ECG aVL", "ECG aVF",
"ECG V1", "ECG V2", "ECG V3", "ECG V4", "ECG V5" and "ECG V6", "SpO
"SpO
".
2
Wave 3: print the third channel of waveform.
⚫
Note
The setting for Wave 1, Wave 2 and Wave 3 should be different (eg. Wave 2 and Wave
3 can not be set as "ECG III" at the same time). Only 2 ECG waveforms at most can
be printed simultaneously.
Print ECG Grids: Switches grids on printed ECG waveform. Select it means the 5x5(mm
⚫
background prints when printing the ECG waveform.
Timer: Switches for the print timer. The default is "OFF". If Print is ON, rotate navigation knob to
⚫
set on the Timer to enable timed print, and set the value of printing intervals in the cycle category.
The setting range is 5~480 minutes; the step is 5 minutes. The default is 60 minutes.
CSM: Select the CSM parameter, options are "EEG" and "Trend" that means the corresponding
⚫
electroencephalogram and graphic trend will be printed respectively.
3.9 Operating Modes
3.9.1 Real-Time Mode
The "Real-Time" shows the normal work mode with real-time waveform and data.
3.9.2 Demo Mode
The "Demo" mode shows the demo working mode with simulated waveform and data. In the "Demo" mode,
all the signals and data are generated by the device to demonstrate and test the monitor (the alarm
function can be checked in this mode).