When changing the sampling tube, it is suggested to choose the default sampling tube
with a dehumidifying function. The sampling tube without dehumidifying function may be
easily blocked by excessive moisture. (Use life: ordinary sampling tube: 6-12 hours; the
sampling tube with dehumidifying function: about 120 hours.)
If the measurement shows an abnormality caused by sampling tube block, replace the
sampling tube block.
The total length of the sampling tube and extending the airway tube shouldn't be longer
than 3 meters. If the tube is too long, measurement errors may occur. If using T connector
sampling cannula kits, insert the sampling tube with the tubes upward to avoid the effects
of excessive moisture.
Altitudes vary for different locations. Set the Barometric Pressure setting value to the
ambient barometric pressure for the location of use.
Use the manufacturer-approved accessories only.
While using the CO
endotracheal tube or a damaged CO
These readings include flow, volume, pressure, and other respiratory parameters.
When stopping the CO
monitor.
Connect the gas outlet to the scavenging system when using sidestream CO
measure CO
.
2
Note
Disposal of the CO
In the presence of electromagnetic devices (i.e., electrocautery), patient monitoring may be
interrupted due to electromagnetic interference. Electromagnetic fields up to 20 V/m will
not adversely affect system performance.
Nitrous oxide, elevated levels of oxygen, helium, and halogenated hydrocarbons can
influence the CO
Excessive moisture in the CO
sensor, a system leak, that may be caused by an uncuffed
2
2
monitor, please disconnect the CO
2
Sensor and its accessories should comply with national and local.
2
measurement.
2
may affect the accuracy of the flow measurement.
2
sensor may significantly affect flow-related readings.
89
sensor from the patient
2
module to
2