The Patient Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the Patient Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 V in ISM and
amateur radio bands
between 0,15 MHz
and 80 MHz
3 V/m
Radiated RF
80 MHz to 2.7 GHz
IEC 61000-4-3
385MHz-5785MHz
Test specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects, and people.
Compliance level
3V
150 kHz to 80 MHz
6 V in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
3 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
166
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Patient Monitor including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
b
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
b