Select "Alarm"→"Others".
2.
In "Others", select "Alarm Volume". Its setting range is "1 - 10" in standard configuration, and the
setting range is "0-10" in spot check configuration, the step is 1. The default is 5. "10" is the maximum
volume.
Caution
When the alarm sound is switched off, the monitor will give no audible alarm tones even if
a new alarm occurs. Therefore, the user should be very careful about whether to switch off
the alarm sound or not.
Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm
volume to a low level may result in a hazard to the patient. Always keep the patient under
close surveillance.
Switching off alarms may result in a hazard to the patient, please be very careful.
If multiple alarms signal at the same time, the monitor shows the high priority alarm with an
audible alarm and alarm lamp. All alarm signals information, including message
descriptions and numerical highlight show respectively and simultaneously on the screen.
Users should not change the alarm volume lower than the factory default setting if close
and constant attention cannot be paid to the patient. Otherwise, the negligence of the
alarm event might cause harm to the patient.
During the alarm reset, any new alarm event can activate the audible alarm again, and the
audible alarm function resumes a normal state.
6.4 Understanding the Alarm Setting
Select the "Alarm" button on Statusbar, and select parameter (ECG, SpO
set alarm limits, alarm switches, high/low alarm limits for each parameter, as shown in the below figure.
Alarm setting window
48
, NIBP, TEMP). You can review and
2