All On: select it to enable alarm function for all parameters. Password is required for this
⚫
operation.
All Off: select it to disable alarm function for all parameters. Password is required for this
⚫
operation, and it's not recommended to disable the alarm function completely.
ARR: to set the alarm priority level for 5 arrhythmia (ARR) events, which include V-Fib/V-Tach,
⚫
Asystole, Bradycardia, Tachycardia, and SV Tachy. There are three priority alarm levels for option:
high, medium and low. The user can also enable or disable the alarm switch for these 5 ARR
events.
Others: to set the alarm volume and timeout setting for Apnea alarm.
⚫
Note
If the parameter alarm switch is set to "OFF", then icon "
corresponding parameter panel. If the parameter panel has two or more related
parameters, once one alarm switch for these related parameters set as "OFF", icon "
will be displayed on its parameter panel. For example, there are SYS, DIA and MAP for
NIBP parameter panel, if the alarm switch for SYS is set as "OFF", then icon "
displayed on the NIBP panel.
The high alarm limit for SpO
For the monitor configured with IBP, CO
alarm limits, and alarm level for these parameters can be set here.
Alarm settings are saved when power off, which means the previous settings continue
even the patient monitor is powered off. Repower includes unexpected power failure and
manual reboot.
6.4.1
High and Low Alarm Setting Range
Select "Alarm"→"Normal Parameter".
Parameter
HR (bpm)
S-T(mV)
SpO
(%)
2
PR (bpm)
TEMP1 (℃)
TEMP2 (℃)
TD (℃)
is fixed to be "100" and is nonadjustable.
2
High limit
Adult: (Low limit+1) -300
Pediatric/Neonate: (Low limit+1) -350
(Low limit+0.01) - 2.50
100
(Low limit+1) -300
(Low limit+0.1) -60.0
(Low limit+0.1) -60.0
" will be displayed on the
, CSM monitoring, the alarm switch, high/low
2
Setting range
0.0-5.0
49
" will be
Low limit
0 - (High limit-1)
-2.5 - (High limit-0.01)
0 - (High limit-1)
0 - (High limit-1)
0 - (High limit-0.1)
0 - (High limit-0.1)
"