Guidance and manufacturer's declaration-electromagnetic emission-
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user
of equipment should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage
fluctuations/flicker
emissions
IEC61000-3-3
Note
The EMISSIONS characteristics of the monitor make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required), the monitor might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures,
such as relocation or re-orienting the monitor.
If the electromagnetic field strength in the location where monitor is used within 1.5 km
from the AM, PM, TV broadcast exceeds the applicable RF compliance level (listed in
Table 3), the monitor should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating Monitor.
for all EQUIPMENT AND SYSTEMS
Compliance
The equipment uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The equipment is suitable for use in all establishments other
Class A
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used
for domestic purposes.
Class A
Complies
Table 1
Electromagnetic environment - guidance
164