NIBP measurement once every 30 minutes, and repeat 5 times); Lastly, entering into Phase E
(making NIBP measurement once every 60 minutes, and repeat once only). In 30 minutes, if the
NIBP measurement is less than 6 times, and NIBP measurement mode is not changed, then the
monitor starts to make NIBP measurement from Phase A to E automatically.
Caution
STAT mode can only be used for an adult.
Note:
When changing the patient, the default NIBP measuring mode is "Manual".
For "Adult", if the NIBP measuring mode is set as "STAT", then the monitor does not save
this setting when you shut down the monitor. When re-starting the monitor, the NIBP
measuring mode for Adults is "Manual".
For all patient types, if the NIBP measuring mode is set as "Manual", or "Auto", or
"Customized Multi-cycle", the monitor saves this setting when you shut down the monitor.
Cycle
it can be set only in the "AUTO" mode. Cycle means the time interval between measurements when the
measuring mode is set to Auto. The Cycle options are: 1min, 2min, 3min, 4min, 5min, 6min, 7min, 8min,
9min, 10min, 15min, 20min, 25min, 30min, 35min, 40min, 45min, 50min, 55min, 1h, 1.5h, 2h, 2.5h, 3h,
3.5h, 4h, 4.5h, 5h, 5.5h, 6h, 6.5h, 7h, 7.5h and 8h. That's, the range is from 1 minute to 10 minutes, step is 1
minute; for range from 10minutes-1hour, step is 5 minutes; for range from 1 hour-8 hours, step is 0.5 hour.
When in AUTO mode and this item is set, the operator must manually press NIBP measure key "
the first time, then the device starts time counting down. The device starts taking the next blood pressure
measurement automatically after finishing counting down.
Unit
the pressure unit. "mmHg" and "kPa" can be selected. 1 kPa =7.5 mmHg.
Init. Pressure
Initial cuff pressure. The range is different depending on patient type.
for adult: 80, 100, 120, 140, 150, 160, 180, 200, 220, 240, 260, 280 mmHg;
for pediatric : 80, 100, 120, 140, 160, 180, 200, 210 mmHg;
for neonate: 60-140 mmHg.
Note
if your monitor is configured with SunTech NIBP module, then the initial cuff pressure range
for adult is 120-200mmHg, the initial cuff pressure range for pediatric is 80- 200 mmHg, the
initial cuff pressure range for neonate is 60 – 80 mmHg.
PARM
to enter into NIBP (SYS, DIA, MAP,) alarm setting. See Section 6.4 "Understanding the Alarm Setting".
73
" for