correctly.
3.
Position the airway adapter in the patient's respiratory circuit (as close to the patient as possible)
between the endotracheal tube and the ventilator circuit. Next, turn on the CO
Screen and then wait 2 minutes for the sensor warm-up.
Note
Always position the sensor with the adapter in an upright position to avoid a collection of
fluids on the windows of the adapter. Large concentrations of fluids at this point obstruct
gas analysis.
13.4 Measurement Limitations
The following factors may influence the accuracy of measurement:
Leaks or internal venting of the sampled gas
Mechanical shock
Cyclic pressure up to 10 kPa (100 cmH
Other sources of interference, if any.
13.5 Troubleshooting the Sidestream CO
When the sampling system of the sidestream CO
kinked. If not, remove it from the water trap. If the monitor gives a message indicating the airway still works
incorrectly, it indicates that the water trap must is blocked, and you should replace it with a new one.
Otherwise, you can determine that the sampling line is blocked. Replace with a new sampling line.
O)
2
Sampling System
2
module works incorrectly, check if the sampling line is
2
92
switch at CO
Setup
2
2