Respiration Panel
Label and unit of respiration
Respiratory Sync mark
The high and low alarm
limit setting of RR
"RR": the label of Respiration. "rpm" is the unit of the Respiration Rate (respiration per minute).
The big font "14" is the value of Respiration rate.
"
": Breath symbol. The blinking frequency is the same as the Respiration Rate.
"
": the high and low alarm limit setting for Respiration Rate.
8.4 Changing RESP Settings
Select "Menu" → "RESP" to enter into RESP related setting.
Gain: Respiration amplification/gain times, 4 options: X1/2, X1, X2, X4. The default is X1 for adult
and pediatric patients, and X2 for neonate patient.
X1/2: half-scale size of the base gain
X1: waveform scale with base gain
X2: twice scale size of the base gain
X4: four times scale size of the base gain
Speed: Respiration waveform sweep speed of, 2 options: 6.25mm/s and 12.5 mm/s. The default is
12.5 mm/s.
Apnea: The timeout setting for apnea alarm (in second). It can be set as any number from 5 to 120
seconds, the step is 1 second. Icon "
Panel; When the device has not detected any breathing signal for the specified time, the "Apnea"
alarm prompts on, and alarm sound will be activated. If it is set as OFF, the icon "
displays on lower left corner of the Respiration Panel.
Note
When "Wires" is set as "10-leadwire" in ECG related settings, then "Apnea" can be set
as off, 10, 15, 20, 25, 30, 35 and 40.
Source: Respiration signal source. This item is fixed to be "CO
selected. Otherwise, the source is obtained by measuring the thoracic impedance via ECG
electrodes, the signal source can come from the ECG electrodes defined by "ECG Lead I" and "ECG
Lead II".
PARM: to enter into RR alarm setting. See Section 6.4 "Understanding the Alarm Setting".
Zero: press it to perform CO
Default: resume to the factory default value.
: the apnea alarm status in RESP alarm setting, refer to the following Section for details.
zeroing.
2
" displays on the lower left corner of Respiration
" if the CO
2
65
Respiration rate
"
monitoring function is
2