Do not use the monitor to those patients who have a severe hemorrhagic tendency or who
have sickle cell disease, for they may develop partial bleeding when this monitor is used
to take the blood pressure measurement.
All the connecting cables and tubes of the applying parts should be kept away from the
patient's neck to prevent any possible suffocation of the patient.
To prevent the risk of a short circuit and to ensure the ECG signal quality, the device must
be properly grounded.
The device should be considered an early warning device as a trend towards patient
deoxygenation is indicated; blood samples should be analyzed by a laboratory
CO-oximeter to understand the patient's condition completely.
Although biocompatibility tests have been performed on all the applied parts, some
exceptional allergic patients may still have anaphylaxis. Do NOT apply to those who have
anaphylaxis.
Please peruse the pertinent content regarding clinical restrictions and contraindications.
The clinical operator should regularly test the device and accessories. A visual and
auditory alarm signal can be checked by disconnecting accessories or by setting it in the
Demo mode to simulate an alarm event.
Do not allow service or maintenance on the device while being used on a patient.
1.1.2 Cautions
DO not replace any of the components in the device with substitute parts. The substitution
of a component different from that supplied by the manufacturer might result in
measurement error. If necessary, please use the components provided by the
manufacturer or those that are of the same model and standards as the accessories along
with the monitor, which are provided by the same factory. Otherwise, there may be
adverse effects concerning safety and biocompatibility. No modification of this device is
allowed.
Repeatedly used accessories should be cleaned thoroughly before use on another patient.
Please refer to the related chapter for the maintenance method.
If the monitor accidentally falls, please do NOT operate it before testing its safety and
technical indexes, with positive testing results obtained.
Do not immerse the monitor or its accessories in liquid to clean.
The system might not meet its performance specifications if stored or used outside the
manufacturer's specified temperature and humidity ranges.
1.1.3 Notes
All combinations of equipment must comply with the standard IEC 60601-1.
The device uses a mains plug as a means of electrical isolation from the mains power. DO
2