19.1 System Information
Select "Menu" → "About...", then the "System Info" window can be entered, which including release
version, software version, hardware version, power version and product ID.
Product ID: set by the manufacturer for traceability.
Embedded Module: other extended modules except the built-in module (ECG/ SpO
such as CSM and CO
module.
2
The current installed module information will also be displayed, for example, "Plug-and-Play modules 1:
CSM-IoC 2: EtCO
".
2
19.2 Nurse Call Settings
Output level: the output level from nurse call connector, "High level" and "Low level" for option.
Duration: two output modes for optional: "Continuous" and "Pulse". The continuous mode of
output means the nurse call signal will keep until the alarm disappears. The pulse mode means
the output nurse call signal lasts for one second.
Trigger Condition: 3 kinds of alarm sources can trig the nurse call: high level alarm, medium level
alarm and low level alarm.
Caution
Do not rely exclusively on the nurse call system for alarm notification. Remember that the
most reliable alarm notification combines audible and visual alarm indications with the
patient's clinical condition.