Your device should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your
place, the device should be cleaned more frequently. Before cleaning the equipment, consult your hospital's
regulations for cleaning the device.
Recommended cleaning agents are:
Sodium hypochlorite bleach (diluted)
⚫
Hydrogen peroxide (3%)
⚫
75% Ethanol
⚫
70% Isopropanol
⚫
To clean your device, follow these rules:
Switch off the monitor and disconnect the power cable before cleaning.
⚫
➢
Kept the monitor from dust.
➢
It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only
non-corrosive cleanser such as clear water is permitted.
➢
Wipe the surface of the monitor and transducers with an Ethanol impregnated wipe, and dry it
with a dry and clean wipe or simply air-dry.
➢
This monitor can be disinfected. Prior to disinfection, clean the monitor.
Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
⚫
Clean the exterior of the connector only.
⚫
➢
Dilute the cleanser.
➢
Do not use abrasive materials.
➢
Do not let any liquid flow into the shell or any parts of the monitor.
➢
Do not let cleanser and disinfectant stay on its surface.
➢
Do not perform high-pressure sterilization to the monitor.
➢
Do not put any parts of the monitor or its accessories in liquid.
➢
Do not pour disinfectant on its surface while disinfecting.
➢
If the monitor accidentally gets wet, thoroughly dry the monitor before use. The rear cover can be
removed by a qualified service technician to verify the absence of water.
➢
Never use this machine in an environment with inflammable gas.
➢
Avoid being hit by lightning. Plug the power cable into an outlet with a grounding wire. Do not use
an outlet in poor condition. If possible, use a power supply system with a regulator.
➢
It must be used in a clean environment protected against shock. Keep it away from corrosive
substances, explosive substances, high temperature, and dampness.
➢
If it is installed in a cabinet, make sure the installation allows for adequate ventilation, and easy
maintenance, observation, and operation.
21.1 Disinfecting the Device and Accessories
Disinfection may cause damage to the device and is therefore not recommended for this monitor unless
otherwise indicated in your hospital's servicing schedule. Cleaning device before disinfecting is
recommended.
The recommended disinfects include:
Chapter 21 Cleaning and Disinfection
132