10.6 Changing SpO
Select "Menu" → "SpO
SE Threshold (%): set the threshold of oxygen desaturation event (%), the value range is from 1 to
⚫
12, step is 1. The default is 3(%).
Beep Volume: to set the pulse beep sound volume. Its setting range is "0 - 10" level. The default is
⚫
5. If you set the beep volume as "0", it means the pulse beep sound is off. It's not recommended
to set it as "0".
PARM: to enter into SpO2 alarm setting. See Section 6.4 "Understanding the Alarm Setting".
⚫
The conditions to trigger SpO
Condition 1:
If the value difference between current SpO
the set threshold, and it keeps this value for at least 8 seconds, then an SE is generated.
Condition 2:
When the current SpO
value and the average value in the previous second is larger than the set threshold, then a SE is generated
too.
10.1 Nellcor SpO
Note
The SpO
monitoring with Nellcor module is not CE-certified.
2
If your monitor is configured with Nellcor SpO
window is as shown in below figure.
SE Threshold(%): set the threshold of oxygen desaturation event (%), the value range is from 1 to
⚫
12, step is 1. The default is 3(%).
SatSeconds: SatSeconds sensitivity setting, to set the max buffering time of activating SpO
⚫
Options are "0", "10", "20", "50" and "100". If you set SatSeconds as "0", it means to turn off the
SatSeconds alarm management function. The over-limit points multiplied over-limit seconds
makes SatSeconds.
SpO
panel for Nellcor SpO
2
and PR Settings
2
" to enter into the SpO
2
Event (SE):
2
2
value is between 90% and 100%, and the value difference between current SpO
2
Module
2
module
2
related setting.
2
value and the average value of the last 1 minute is larger than
Module, then the SpO
2
79
2
related setting
2
alarm.
2