Gima K10 Manuel D'utilisation page 91

Moniteur multiparamètres

Publicité

Les langues disponibles

Les langues disponibles

10.6 Changing SpO
Select "Menu" → "SpO
SE Threshold (%): set the threshold of oxygen desaturation event (%), the value range is from 1 to
12, step is 1. The default is 3(%).
Beep Volume: to set the pulse beep sound volume. Its setting range is "0 - 10" level. The default is
5. If you set the beep volume as "0", it means the pulse beep sound is off. It's not recommended
to set it as "0".
PARM: to enter into SpO2 alarm setting. See Section 6.4 "Understanding the Alarm Setting".
The conditions to trigger SpO
Condition 1:
If the value difference between current SpO
the set threshold, and it keeps this value for at least 8 seconds, then an SE is generated.
Condition 2:
When the current SpO
value and the average value in the previous second is larger than the set threshold, then a SE is generated
too.
10.1 Nellcor SpO
Note
The SpO
monitoring with Nellcor module is not CE-certified.
2
If your monitor is configured with Nellcor SpO
window is as shown in below figure.
SE Threshold(%): set the threshold of oxygen desaturation event (%), the value range is from 1 to
12, step is 1. The default is 3(%).
SatSeconds: SatSeconds sensitivity setting, to set the max buffering time of activating SpO
Options are "0", "10", "20", "50" and "100". If you set SatSeconds as "0", it means to turn off the
SatSeconds alarm management function. The over-limit points multiplied over-limit seconds
makes SatSeconds.
SpO
panel for Nellcor SpO
2
and PR Settings
2
" to enter into the SpO
2
Event (SE):
2
2
value is between 90% and 100%, and the value difference between current SpO
2
Module
2
module
2
related setting.
2
value and the average value of the last 1 minute is larger than
Module, then the SpO
2
79
2
related setting
2
alarm.
2