CSM records can be reviewed by "Trends" and "Events".
The default "CSM Records" reviewing window is by "Trends". All CSM records (including CSI, SQI, EMG% and
BS%) and EEG of the current patient can be reviewed.
In the window as shown in above Figure, the positioning measurement of the CSM trend graph can be
entered. The measured value that the cursor line indicates will be displayed on the lower right side, for
example, "
lower middle side, for example, "
SQI, EMG%, BS% and EEG.
"
": The event mark. The number is the event No., the letter is the event type. The event type
⚫
description is as below:
➢
G: General
➢
U: Intubation
➢
S: Surgery
➢
N: Note
➢
M: Movement
➢
I: Induction
➢
C: Maintenance
➢
J: Injection
➢
E: End Maintenance
Sampling Interval: sample time interval of the trend graph, 6 options: 1 second, 5 seconds, 10
⚫
seconds, 30 seconds, 1 minute and 5 minutes.
The current event records can be reviewed by "Events"
CSM record---by Trends
", and the EEG waveform will be displayed on the
". That's the positioning measurement for CSI,
104
CSM trend graph recall