Measurement accuracy
24.5 SpO
2
SpO
from KRK Module
2
Compliant standards
Measurement range
Measurement accuracy
Sensor
Data update cycle
Perfusion index (PI) measurement
range
Alarm accuracy
SpO
from Nellcor Module
2
Compliant standards
Measurement range
Measurement accuracy
Sensor
Data update cycle
Perfusion index (PI) measurement
range
Alarm accuracy
Accuracy of SpO
Sensors
2
No.
Model/Part No.
1
Sensor: DS100A
DIA
mmHg
kPa
Average deviation ≤±5mmHg (±0.67 kPa)
Standard deviation ≤8mmHg (1.067 kPa)
ISO 80601-2-61: 2011
0%-100%
70%-100%: ±2%
⚫
50%-69%: ±3%
⚫
0%-49%: not defined
⚫
Wavelength: Red light: 660nm, Infrared light: 905nm
Maximum optical output power: ≤2mW
≤8s
0.1%-20%, the measurement accuracy is not defined
70%-100%: ±1%
ISO 80601-2-61: 2011
0%-100%
Adult/Pediatric: 70%-100%: ±2%
⚫
Neonate: 70%-100%: ±3%
⚫
0%-69%: not defined
⚫
Wavelength: Red light: 660nm, Infrared light: 900nm
Maximum optical output power: ≤15mW
≤8s
0.1%-20%, the measurement accuracy is not defined
70%-100%: ±1%
147
20-200
20-160
2.7-26.7
2.7-21.3
Applicable patient
Adult/Pediatric
20-100
2.7-13.3
Accuracy
Arms (%)
±2