7.4 Understanding the ECG Display
Your display may be configured to look slightly different.
ECG waveform
"ECG": parameter label.
"III": ECG lead. III means ECG lead III.
"25mm/s": ECG waveform sweeping speed, unit is "mm/s".
"X1": ECG waveform gain. "X1" means the waveform scale with base gain.
"MONI": ECG filtering mode. There are three types: diagnose, monitoring, and operation.
"S-T+0.02mV": S-T segment value, here indicates the value is 0.02mV.
Heart Rate Panel
Heart rate label and unit
Heart beat symbol
High and low alarm limit
setting for Heart rate
Arrhythmia symbol
"HR": heart rate. The 64 on the right is the heart rate measured.
"bpm": the unit of heart rate, it means "beats per minute".
"
": the heart-beat symbol, blinks corresponding to the R wave of ECG waveform.
"180/40": high and low alarm limit setting for heart rate.
"
": Arrhythmia symbol. If the ARR function is set as ON, then this symbol will be displayed.