Verification
To verify the NIBP pneumatic system, it is used to verify the pressure accuracy and check the air leakage of
the pneumatic system, which should be conducted by technicians in the specified environment. It includes
"Verification A", "Verification B," and "Leakage Check".
Verification A & Verification B
relevant button to start Verification A or B. When completed, it is necessary to stop the
verification manually by pressing the "Stop" button. The system stops the verification when
the "Exit" button pressed.
Leakage Check: it is used by technicians to perform a leakage inspection for NIBP pneumatic
system.
Note
Verification A: The monitor will auto inflate to the preset pressure value (depending on the
patient type) by the pump, then it closes the deflating valve. Comparing the pressure value
on the device with that on standard pressure manometer, and verify whether the pressure
accuracy exceeds the rated tolerance range.
The auto inflating pressure value for Adult: 250 mmHg (33.3 kPa)
The auto inflating pressure value for Pediatric: 200 mmHg (26.7 kPa)
The auto inflating pressure value for Neonate: 125 mmHg (16.7kPa)
Verification B: The monitor closes the valve; the pressure should be inflated manually.
Comparing the pressure value on the device with that on standard pressure manometer,
and verify whether the pressure accuracy exceeds the rated tolerance range.
If the verification window pops up during verification, you can press the "Close" button to
exit from it only, but DO NOT from by Verification B by pressing "Display View Key".
The 2 kinds of verification of the pressure accuracy. Press the
: