Incompatible components can result in degraded performance.
Operation Procedures for thermal temperature transducer
1.
Securely attach the transducer to the patient;
2.
Connect the cable to the TEMP probe connector marked "TEMP" in the panel.
3.
Check that the menu setting is matching the used temperature sensor type.
4.
Check that the alarm settings are appropriate for this patient.
Note
When unplugging the probe, be sure to hold the head of the connector and pull it out.
11.4 Understanding the TEMP Display
1
2
"TEMP1 ℃": Temperature 1 mark and unit. Temperature unit: ℃ or ℉。
1
2
"36.5": displays the temperature measured at temperature channel 1.
3
"TEMP 2": Temperature channel 2.
4
"37.0": displays the temperature measured at temperature channel 2.
5
"TD 0.5": Temperature difference, namely the difference between temperature 1 and temperature 2.
11.5 Changing TEMP Settings
Select "Menu" → "TEMP" to enter into TEMP related setting.
Temp Unit:Temperature unit. The default is ℃ (Celsius) and can be set as ℉ (Fahrenheit).
⚫
The default is "℃".
Probe type: KRK can be selected. If the user type and the set type are different, then the
⚫
measured value is invalid.
TD: the absolute temperature difference. When the temperature difference is higher than the
⚫
preset value, then the device activates an alarm. If there is only one temperature, the TD shows
"――".
PARM: to enter into RESP alarm setting. See Section 6.4 "Understanding the Alarm Setting".
⚫
3
4
82
5