Select "Menu"→"Patient Info.".
In the Patient Info. Window, the operator can add, edit, delete, and apply a patient's document as the
current document for monitoring.
4.1 Apply an existed patient's document
Select "Menu"→"Patient Info."→"Apply".
If there is an existing patient document for the monitored patient, it is unnecessary to create a new
document for the patient. Press the "Apply" button to make the existed document as the "Current Patient's
Document".
4.2 Create a new patient's document
Select "Menu,"→"Patient Info."→"Create".
To input characters at an edit-box, press the Navigation Knob (same as "Enter" key), then the soft keyboard
pops up immediately.
The following items need to input:
PID: The Patient ID.
Group: 3 types of patients: Adult, Pediatric and Neonate. The default is "Adult".
Adult: age > 12 years
Pediatric: 1 month < age ≤ 12 years
Neonate:
First Name: patient first name.
Last Name: patient's last name.
Gender: patient gender.
Important notes: When power on the monitor or the patient type is changed (e.g., the patient type changes
from Adult to Neonate), then the monitor performs initialization. Therefore, please confirm that the patient
type matches the patient to be monitored before monitoring. When "PED" or "NEO" is set to an adult
patient, it is difficult to get a valid NIBP measurement value and easy to make false alarms. It's forbidden to
use the "Adult" mode on a pediatric or neonate patient, for it may cause serious injury.
Chapter 4 Patient Data Management
age ≤ 30 days
33