Gima K10 Manuel D'utilisation page 170

Moniteur multiparamètres

Publicité

Les langues disponibles

Les langues disponibles

Status / Error
Sensor off
Sensor not ready
Zero in progress
Zero failed and
breaths detected
Battery
Status / Error
Low battery
25.2 Common Faults
Check the solutions below before requesting service, if you encounter the problems when using the
monitor or accessories. If the problem persists, contact your service personnel.
Display Troubleshooting
When there is no display on the screen, follow the steps below:
Shut down the device and unplug the power cable.
Use a universal meter to check if the outlet has proper voltage.
Description
the expected range.
When the software system fails
to receive any CO
data for
2
more than 2 seconds, it is
considered that the sensor is
off at this time.
When setting the zero
operation, the sensor status is
not ready. Possible reasons
include:
Breaths detected
The preheating is not
completed, and the sensor
temperature is not stable
The sensor is still in sleep mode
Sensor barometric pressure and
gas compensation settings are
not completed
The status information
returned by the module during
the zero operation.
Breathing is detected within 20
seconds of the zero operation.
Corrective Actions
Charge the battery in time
Corrective Actions
unblocked.
Monitoring application software setting status,
not reported by the module itself, not an error or
fault.
Normal report during zero operation, not error or
fault.
Normal report during zero operation, not error or
fault.
It is recommended to ensure that there is no
breathing gas for at least 20 seconds before the
zeroing operation, and perform the zero
operation under the condition of clear air to avoid
operation failure or zero deviation.