If an alarm limit violation triggers an alarm, the numeric measurement in alarm is highlighted.
6.3.4 Audible Alarm Tones
The alarm tone is distinct from the heartbeat tone, keystroke tone, and pulse tone in frequency.
The alarm tones identify the alarm levels as follows:
High priority alarm: beep+beep+double+beep+pause+beep+beep+double+beep
⚫
Medium priority alarm: triple beep
⚫
Low priority alarm: the single beep
⚫
Caution
When multiple alarms of different levels coincide, the monitor selects the alarm that has
the highest level and gives visual and audible alarm indications accordingly.
46