INTOUCH® CRITICAL CARE BED (CONTINUED)
Recommended separation distances between portable and mobile RF
communications equipment and the InTouch® Critical Care Bed.
The InTouch® Critical Care Bed is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled . The customer or the user of the InTouch® Critical Care Bed can help
prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the InTouch® Critical Care Bed as recommended below,
according to the maximum output power of the communications equipment .
Rated maximum output
power of transmitter
W
0 .01
0 . 1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer .
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
Note 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
Return To Table of Contents
1-76
EMC Information
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d=1 .2√P
0 . 1 2
0 .38
1 .2
3 .8
12
2151-009-005 REV B
m
800 MHz to 2 .5 GHz
d=1 .2√P
0 . 1 2
0 .38
1 .2
3 .8
12
d=2 .3√P
0 .23
0 . 7 3
2 .3
7 .3
23
www.stryker.com