MAIN MENU: SUPPORT SURFACES (CONTINUED)
PositionPRO™ Mattress Specifications
Refer to the PositionPRO™ manual .
Interaction Between InTouch® and
Position PRO™
Siderails Position
The position of the siderails determines the functionality
of the PositionPRO™ mattress . All the therapies, including
Max Inflate, Firmness and Turn Assist, will function if the
siderails are unlocked . If a siderail is unlocked during
the rotation therapy, this therapy will automatically
stop . If two therapies are simultaneously executed and
a siderail is unlocked during these therapies, they will
automatically stop . If one of the siderails is unlocked
during a rotation therapy (ex . rotation and vibration is
being executed simultaneously), both therapies will stop
instantly .
Specific PositionPRO™ Functionality on the
InTouch®
The indicators on the PositionPRO™ menu screen from
the InTouch® and on the PositionPRO™ pendant are
slightly different, but the same functions are available .
When requesting a Turn Assist left or right on the
PositionPRO™ via the InTouch®'s touch screen, a Turn
Assist menu screen will appear (see Figure 10 .3) .
If the InTouch® senses an error with the PositionPRO™,
the Call Maintenance LED will turn on (see Figure 10 .3) .
When activating the InTouch® CPR pedal, the
PositionPRO™ automatically turns on .
When the PositionPRO™ is connected to the InTouch®,
the PositionPRO™ uses the fowler and Trendelenburg
tilt sensors of the InTouch® to sense the angle of the
fowler .
Return To Table of Contents
1-40
Footboard Operation Guide
2151-009-005 REV B
PositionPRO
TM
120:00
Timer
Max Inflate
Left Turn
Firmness (mmHg)
Alarms
14
18
22
26
30
Unit/Rm :
Figure 10.3
PositionPRO
TM
Timer
Start
Right Turn
Matress Turn In Progress
Unit/Rm :
Figure 10.4
PositionPRO
TM
Timer
Start
Left Turn
Matress Turn In Progress
Unit/Rm :
Figure 10.5
Minutes
Stop
Right Turn
Lock All
Call
CPR
Maintenance
Reset
CPR
Oct. 25,2007
12:00 AM
+
120:00
Minutes
-
Back
Oct. 25,2007
12:00 AM
+
120:00
Minutes
-
Back
Oct. 25,2007
12:00 AM
www.stryker.com

Publicité

loading

Ce manuel est également adapté pour:

Intouch 2151Intouch 2152