note: All efforts should be taken to assure that the stent is not underdilated. If the
deployed stent size is still inadequate with respect to vessel diameter, or if full contact
with the vessel wall is not achieved, a larger post-dilatation balloon catheter may be
used to expand the stent further. The balloon should be centered within the stent and
should not extend outside of the stented region.
note: Care must be exercised when crossing a newly deployed stent with an
intravascular ultrasound (IVUS) catheter, a coronary guidewire, or a balloon catheter
to avoid disrupting the stent placement, apposition, geometry, and/or coating. If
recrossing with a guidewire, the stented segment should be recrossed carefully with
a prolapsed tip to avoid dislodging the stent.
6.
Complete angiographic confirmation, remove devices, and close vascular access site
according to standard practice.
In vItro InforMatIon
synerGy™ stent system compliance
table 4. synerGy compliance chart
pressure
atm - kpa
2.25
2.50
8 - 814
---
2.35
9 - 910
2.13
2.42
10 - 1014
2.19
2.48
11 - 1117
2.24
2.54
12 - 1213
2.28
2.59
13 - 1317
2.31
2.63
14 - 1420
2.35
2.67
15 - 1517
2.37
2.70
16 - 1620
2.40
2.73
17 - 1724
2.43
2.76
18 - 1827
2.45*
2.79*
19 - 1924
2.48
2.82
20 - 2027
2.51
2.85
21 - 2130
2.54
2.90
22 - 2227
2.58
2.95
*Rated Burst Pressure. DO NOT EXCEED
Nominal Pressure = 11.0 atm - 1117 kPa
Black (K) ∆E ≤5.0
stent I.D. (mm)
2.75
3.00
3.50
4.00
2.57
2.89
3.30
3.81
2.65
2.96
3.40
3.91
2.72
3.02
3.48
3.98
2.79
3.08
3.55
4.06
2.85
3.13
3.61
4.12
2.89
3.17
3.66
4.17
2.93
3.20
3.70
4.22
2.96
3.24
3.74
4.26
3.00
3.27*
3.79*
4.30*
3.03
3.32
3.83
4.36
3.06*
3.37
3.87
4.42
3.10
3.43
3.93
4.52
3.13
3.49
3.99
---
3.19
---
---
---
3.23
---
---
---
pressure
atm - kpa
2.25
2.50
8 - 814
---
2.51
9 - 910
2.29
2.58
10 - 1014
2.35
2.64
11 - 1117
2.40
2.70
12 - 1213
2.44
2.75
13 - 1317
2.47
2.79
14 - 1420
2.51
2.83
15 - 1517
2.53
2.86
16 - 1620
2.56
2.89
17 - 1724
2.59
2.92
18 - 1827
2.61*
2.95*
19 - 1924
2.64
2.98
20 - 2027
2.67
3.01
21 - 2130
2.70
3.06
22 - 2227
2.74
3.11
*Rated Burst Pressure. DO NOT EXCEED
Nominal Pressure = 11.0 atm - 1117 kPa
warranty
Boston Scientific Corporation (BSC) warrants that reasonable care has been used in the
design and manufacture of this instrument. this warranty is in lieu of and excludes all other
warranties not expressly set forth herein, whether express or implied by operation of law or
otherwise, including, but not limited to, any implied warranties of merchantability or fitness
for a particular purpose. Handling, storage, cleaning and sterilization of this instrument as
well as other factors relating to the patient, diagnosis, treatment, surgical procedures and
other matters beyond BSC's control directly affect the instrument and the results obtained
from its use. BSC's obligation under this warranty is limited to the repair or replacement of
this instrument and BSC shall not be liable for any incidental or consequential loss, damage or
expense directly or indirectly arising from the use of this instrument. BSC neither assumes, nor
authorizes any other person to assume for it, any other or additional liability or responsibility
in connection with this instrument. bsc assumes no liability with respect to instruments
reused, reprocessed or resterilized and makes no warranties, express or implied, including
but not limited to merchantability or fitness for a particular purpose, with respect to such
instruments.
Viagra is a trademark of Pfizer, Inc.
Xience V is a trademark of Abbott Cardiovascular Systems, Inc.
Magnetom Trio and syngo are trademarks of Siemens Aktiengesellschaft.
Intera is a trademark of Koninklijke Philips Electronics NV.
10
stent o.D. (mm)
2.75
3.00
3.50
4.00
2.73
3.05
3.46
3.99
2.81
3.12
3.56
4.09
2.88
3.18
3.64
4.16
2.95
3.24
3.71
4.24
3.01
3.29
3.77
4.30
3.05
3.33
3.82
4.35
3.09
3.36
3.86
4.40
3.12
3.40
3.90
4.44
3.16
3.43*
3.95*
4.48*
3.19
3.48
3.99
4.54
3.22*
3.53
4.03
4.60
3.26
3.59
4.09
4.70
3.29
3.65
4.15
---
3.35
---
---
---
3.39
---
---
---