Pod assembly replacement
Tools required:
•
Zipper pull tool or equivalent
Procedure:
1.
Remove the top cover. See
2.
Unsnap the pillow (H) from the left and right side of the pod assembly
3.
Separate the six Velcro® connectors holding the foot section gel crib to the pod assembly.
4.
Remove the foot section gel crib (F) by lifting upward
5.
Disconnect the quick disconnect fitting (Z)
6.
Remove the two retaining pins (A) from the hose quick connection from the foot box assembly
1-38). Save the pins.
www.stryker.com
Service
Cover replacement, top on page 1-34
(Figure 1-12 on page
Figure 1-12: Gel crib, pillow, and quick disconnect fitting
(Figure 1-12 on page
2971-109-0 01 REV B
for removal.
(Figure 1-12 on page
1-37). Save the pillow.
1-37). Save the gel crib.
Z
1-37).
(Figure 1-13 on page
1-37
E E N N