3 Using the Source/Load Functions
Configuring an Output Turn-On/Turn-Off Sequence
Turn-on and turn-off delays control the turn-on/turn-off timing of the outputs in relation to each other.
You can also synchronize Output On/Off delays across multiple mainframes. Refer to
Output Couple Controls
Set the output voltage and current of the output channels
Refer to
Turning the Unit On
Configure the turn-on turn-off delays
Press the Settings key twice to access the Output On/Off Delays window. Enter the On Delays and Off
Delays for all outputs that will participate in the output on/off delay sequence. Values can range from
0 to 1023 seconds.
All power modules have an internal turn-on delay that applies from the time that the command to turn
on the output is received until the output actually turns on. This turn-on delay is automatically added
to the On Delays values. The turn-on delay does not apply when outputs turn off. To view the delay,
select the Output Coupling button.
Normally, the firmware automatically calculates the delay offset based on the longest minimum turn-
on delay of the installed power modules. However, if you will be excluding some outputs from
participating in an output on/off delay sequence as described Step 3, the delay offset may be different
based on the outputs that you will actually be sequencing. The minimum turn-on delays of the power
modules are documented in the
98
for further information.
and set the output voltage and current values of all sequenced outputs.
Keysight N6700 Modular Power System Family Specifications