Step 2.Connect the wires to the power module.
Connect the load wires to the output connector of the power module as shown. Connect the remote
sense wires to the +S and -S terminals on the sense connector.
The +LS and –LS terminals are ONLY used for local sense connections to the +S and
- S terminals. Do not connect the +LS and –LS terminals in any other way.
Step 3. Finish the installation.
Place any unused cable harnesses in the clip ring located between the power modules and the front
panel. Install the top and bottom covers. Push the covers into place and tighten the thumbscrews.
1. 50 A output
connector
2. Sense
connector
3. Twist leads
4. To load
Ferrite Core Installation - for Keysight N6792A only
Compliance to radio frequency interference (RFI) standards requires a ferrite core to
be installed on the load leads of the load module. The core is supplied with the
module and does not affect the functionality of the module.
1. Locate the core along the length of the cables and pass the load cables through the core once
2. Clip the ferrite core onto the load cables as close as possible to the back of the unit. The
recommended distance from the connector to the core is 4 cm.
3. Attach the tywrap to the load wires to prevent the core from sliding away from the back of the unit.
Keysight N6705C User's Guide