Télécharger Imprimer la page

Keysight N6705C Mode D'emploi page 52

Masquer les pouces Voir aussi pour N6705C:

Publicité

Les langues disponibles

Les langues disponibles

2 Installation
Installing the Power Analyzer
Power Module Installation
High-Current Output Connections
Ferrite Core Installation - for Keysight N6792A only
Bench Installation
Rack Installation
Redundant Ground for 400 Hz Operation
Power Module Installation
The information in this section applies if you have purchased an N6705 mainframe
without the power modules installed, or if you are adding a power module to the
mainframe.
Equipment Damage Turn the mainframe off and disconnect the power cord before
installing or removing power modules. Observe all standard electrostatic discharge
precautions before handling electronic components.
Tools required: T10 Torx driver; Small flat-blade screwdriver; 5.5 mm hex wrench
Step 1. Remove the top and bottom covers.
Loosen the thumb-screws to remove the covers. Slide the covers back. Turn the unit upside-down to
remove the bottom cover.
Step 2. Place the power modules in the mainframe.
Align the power module over the pins and gently push it down onto the connector. Install the screws
at each end of the power module. The torque specification for the screws is 9 inch-pounds.
52
Keysight N6705C User's Guide

Publicité

loading