6 Advanced Source, Measurement, and Control
The frequency parameter sets the pole frequency associated with a voltage bandwidth range. The
power-on default for each range is the minimum frequency. This is optimized for no overshoot with the
worst-case load capacitance for that range. If the load capacitance is not worst-case, or if some
output overshoot is acceptable, the output voltage transition times can be reduced by increasing the
frequency limit.
From the remote interface:
The bandwidth is selected using the following SCPI command:
VOLT:BWID <LOW|HIGH1|HIGH2|HIGH3>, (@1)
When queried, the bandwidth that was selected is returned.
To select a different frequency limit, use the following SCPI command:
VOLT:BWID:LEV <LOW|HIGH1|HIGH2|HIGH3>, <frequency>, (@1)
When queried, the selected frequency is returned.
Keysight N679xA Load Module Operation
The operating modes of operation for Keysight N679xA Load modules
are: current
priority, voltage priority, power priority, and resistance priority. When programmed to a mode, a
module remains in that mode until the mode is changed or until a fault condition, such as an over-
power or over-temperature occurs.
Current Priority Mode
In this mode, the load module will sink a current in accordance with the programmed value regardless
of the input voltage.