4 Using the Measurement Functions
SENS:DLOG:FUNC:MINM ON
To specify a data log of 1000 seconds on all enabled outputs:
SENS:DLOG:TIME 1000
To specify a sample period of 50 milliseconds between data samples on all enabled outputs:
SENS:DLOG:PER .05
Data Logger Ranges
For outputs that have multiple measurement ranges, you can select a lower range for better
measurement resolution. Data Logger range settings are independent of the Meter View and Scope
range settings.
From the front panel:
In the Data Log Ranges area, select the desired lower measurement range from the Voltage or
Current dropdown menus.
Some models have a high measurement range labeled CComp On, which is selected by default. The
CComp On range compensates output current measurements during voltage transients. For more
information, see
Dynamic Current Measurement
On Keysight N679xA Load modules
are coupled. Therefore the data logger measurement ranges are not settable (grayed
out) in the Data Logger Properties window.
Seamless Measurements
Seamless measurement autoranging for both voltage and current is available on models N678xA SMU
and with Option SMR.
with no data lost across ranges.
Seamless ranging does not include the 10 μA range, which must be selected manually. Select Auto to
enable seamless measurement ranging.
160
Control.
, the source and measurement ranges
This enables a wide dynamic measurement range
Keysight N6705C User's Guide