5 Using the System Functions
Configuring the Remote Interfaces
GPIB Configuration
USB Configuration
LAN Configuration
Modifying the LAN Settings
Using the Web Interface
Using Sockets
Using Telnet
Securing the LAN
This Keysight N6705C DC Power Analyzer supports remote interface communication over three
interfaces: GPIB, USB, and LAN. All three interfaces are live at power on. For information on
connecting the interfaces, refer to
Interface
Connections.
The front panel IO indicator comes on whenever there is activity on the remote interfaces. The front
panel Lan indicator comes on when the LAN port is connected and configured.
Ethernet connection monitoring is also provided. With Ethernet connection monitoring, the
instrument's LAN port is continually monitored, and automatically reconfigured when the instrument is
unplugged for a minimum of 20 seconds and then reconnected to a network
GPIB Configuration
Each device on the GPIB (IEEE-488) interface must have a unique whole number address between 0
and 30. The instrument ships with the address set to 5. Your computer's GPIB interface card address
must not conflict with any instrument on the interface bus. This setting is non-volatile; it will not be
changed by power cycling or *RST.
To access the GPIB parameters from the front panel, press the Menu key, select Utilities, then I/O
Configuration, then GPIB/USB. Check or uncheck Enable GPIB to enable or disable the GPIB. Use the
numeric keys to enter a value in the Address field. Press Enter to enter the value.
190
Keysight N6705C User's Guide