Symbol/Field
1 m1/m2 points
2 Delta
3 m2
4 m1
5 Min
6 Avg
7 V p-p
8 Trigger Mode
Ah (if selected)
Wh (if selected)
Using the Waveform Display Knobs
Keysight N6705C User's Guide
Description
Shows where the measurement markers intersect the selected waveform. Data values at the bot-
tom of the display are referenced to the intersect locations of the markers. Calculations are based
on the data points in between the intersect locations.
Indicates the delta or absolute difference between the markers in units (volts, amps, or watts) and
in time (seconds).
Indicates the m2 marker value in volts, amps, or watts at the intersection point. Also indicates the
distance in time that the m2 marker is in relation to the present trigger position.
Indicates the m1 marker value in volts, amps, or watts at the intersection point. Also indicates the
distance in time of the m1 marker in relation to the present trigger position.
Indicates the minimum data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Also indicates the distance in time of the minimum value in relation to the
present trigger position.
Calculates the average data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Time indicates the time between markers over which the average value is
calculated.
Indicates the maximum data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Also indicates the distance in time of the maximum value in relation to the
present trigger position.
Calculates the difference between the maximum and minimum values. Time information is not
valid for calculated p-p values.
Calculates the Amp-hours between the marker locations. To view the Amp hours, you may need
to unselect one of the other measurements in the Datalogger Marker Properties window. Only 5
measurements may be displayed at a time.
Calculates the Watt-hours between the marker locations. To view the Watt hours, you may need to
unselect one of the other measurements in the Datalogger Marker Properties window. Only 5
measurements may be displayed at a time.
4 Using the Measurement Functions