The I
and I
parameters shown apply
0
1
to the Current Arb type.
The V
, V
, P
, P
, R
, and R
0
1
0
1
0
parameters apply to the respective
Voltage, Power and Resistance Arb
types.
Parameter
Description
Start Setting (I 0 , V 0 , P 0 , or
The setting before the staircase.
R 0 )
End Setting (I 1 , V 1 , P 1 , or
The setting after the final step. The difference between the start and end setting is divided equally between steps.
R 1 )
Delay (T 0 )
The delay after the trigger is received but before the staircase starts.
Step Time (T 1 )
The time to complete all staircase steps.
End Time (T 2 )
The time the output remains at the end setting after the staircase completes.
# of Steps
The total number of staircase steps.
Step 3. Configure the final steps common to all Arbs.
Refer to
Configuring Parameters Common to all Arbs
From the remote interface:
Select the Arb Type and Shape:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP STA,(@1)
Enter the values for the level before and after a voltage staircase:
ARB:VOLT:STA:STAR 0,(@1)
ARB:VOLT:STA:END 5,(@1)
Enter the time before the staircase, the stair-time of the ramp, and the time after the staircase:
ARB:VOLT:STA:STAR:TIM 0.2,(@1)
ARB:VOLT:STA:TIM 0.2,(@1)
ARB:VOLT:STA:END:TIM 0.2,(@1)
Enter the total number of staircase steps:
ARB:VOLT:STA:NST 3,(@1)
Keysight N6705C User's Guide
1
3 Using the Source/Load Functions
at the end of this section.