Step 2 – Configure the output turn-on sequence:
Configure the output turn-on sequence as described under
Note that you will only need to configure the Turn-on delays, not the turn off delays. The turn on
delays for the output channels are as follows:
Output 1: 10 ms
l
Output 2: 20 ms
l
Output 3: 30 ms
l
Output 4: 40 ms
l
Step 3 – Configure the Scope View traces:
Check V1 through V4
l
Uncheck I1 through I4
l
Use the Vertical Volts/Div. knob to set V1 through V4 to 10 V/Div.
l
Use the Offset knob to move the four traces so that they are separated by at least one space on the
l
vertical grid.
Use the Horizontal Time/Div. knob to set the timebase to 20 ms.
l
Step 4 – Configure the Scope properties:
Press the Properties key to configure the scope properties as follows:
In the Trigger Source dropdown list, select Output On/Off Key.
l
In the Mode dropdown list, select Single.
l
Select the Horizontal button and make sure that the Horizontal Offset Reference is set to Left.
l
Keysight N6705C User's Guide
4 Using the Measurement Functions
Configuring a Turn-On/Turn-Off
Sequence.