Selecting Knob Control allows the Vertical Volts/Div knob on the front panel to automatically select
lower measurement ranges at higher measurement resolutions. For lower measurement resolution a
higher range is automatically selected. The present range is displayed on the lower left corner of the
Scope View window.
Some models have a high measurement range labeled CComp On, which is selected by default. The
CComp On range compensates output current measurements during voltage transients. For more
information, see
Dynamic Current Measurement
On Keysight N679xA Load modules
are coupled. Therefore the scope measurement ranges are not settable (grayed out) in
the Scope Properties window.
Seamless Measurements
For Models N6781A, N6782A, N6785A, and N6786A, you can select seamless voltage and current
measurements, which results in a wide dynamic range with no data being lost across ranges. Select
Auto to enable seamless measurement ranging. Seamless ranging does not include the 10 μA range,
which must be selected manually.
Scope Markers
Select the Markers button to configure the measurements that appear on the bottom of the display in
Marker view. Measurements apply to the portion of the waveform between the two markers. You can
only select a maximum of five measurements to be displayed.
Keysight N6705C User's Guide
4 Using the Measurement Functions
Control.