3 Using the Source/Load Functions
Arb sequences are saved and recalled in instrument state files, but you can also save and load them in
files that are separate from other instrument settings.
Select the Load button to load the file. Select the Save button to save the file. Then provide the
following information:
Parameter
Action
Type
Output1
Path/Filename
Browse
Load
Save
Step 4. Configure the final steps common to all Arbs.
Press the Advanced button to specify what happens when the Arb sequence completes.
Refer to
Configuring Parameters Common to all Arbs
118
Description
The Load or Save action.
The Sequence file type is .seq
When loading, the output on which the sequence will run.
When saving, the output from which the sequence is saved.
The path and filename of the sequence. Internal:\ specifies the instrument's internal memory.
External:\ specifies the Memory port on the front panel.
Specifies where the sequence file is located or where the sequence file will be placed.
Loads the file.
Saves the file.
at the end of this section.
Keysight N6705C User's Guide